Product Features
◆ Double-way high-precision DC current, voltage, resistance signal and thermocouple, thermal
resistance temperature measurement;
◆ 6-bit digital display, resolution of 0.001℃, large screen display window;
◆ Digital temperature control, rapid rise and fall temperature, convenient setting, good temperature
control stability;
◆ The average heat block can be customized;
◆ The horizontal temperature field, vertical temperature field are good;
◆ Furnace body built-in load short circuit, load open circuit, sensor protection and a series of protection functions;
◆ Small volume, light weight, easy to carry, no pollution;
◆ Support for USB / R232, WiFi communication.
Instrument Network Version Function (Added)
◆ The instrument can be linked to the Internet of Things system platform through WiFi with one click;
◆ Verification data automatically uploaded to the cloud, automatic processing, free customers cumbersome operation;
◆ Verification of the original records, reports and certificates are permanently saved in the cloud and can be manually deleted;
◆ Using B/S architecture cloud service, customers can log in anytime and anywhere, support PC and mobile phone login.
Product Name |
YH150 Dry Well Furnace |
YH650 Dry Well Furnace |
YH1200 Dry Well Furnace |
Temperature Range |
-20℃~150℃ |
(50~650)℃ |
(150~1200)℃ |
Accuracy |
±0.15℃ |
±0.5℃ |
±3℃ |
Display Resolution Ratio |
1, 2, 3 digits can be set |
1, 2, 3 digits can be set |
0.1℃ |
Insertion Depth |
165mm |
165mm |
165mm |
Diameter of Uniform Heating Block |
25mm |
30mm |
30mm |
Aperture of Uniform Heating Block |
Φ4; Φ6; Φ8 |
Φ4; Φ6; Φ8; Φ10 |
Φ4; Φ6; Φ8; Φ10 |
Aperture of Uniform Heating Block |
The diameter and aperture of the heating block can be customized according to customer needs |
The diameter and aperture of the heating block can be customized according to customer needs |
The diameter and aperture of the heating block can be customized according to customer needs |
Horizontal Temperature Field |
≤±0.05℃ |
≤±0.05℃ |
≤±0.2℃ |
Vertical
Temperature Field |
Within 50mm from the bottom of the trap, ≤ 0.5 ℃ |
Within 50mm from the bottom of the trap, ≤ 0.5 ℃ |
Within 40mm from the bottom of the trap, ≤ 0.5 ℃ |
Temperature Field Fluctuation |
≤±0.03℃/min |
≤±0.05℃/min |
±0.2℃ |
Power |
800W |
1200W |
1500W |